The OmniPure Sub-Boiling Distillation System is a fully automated unit with all wetted parts composed of inert, high-purity Teflon® PFA and FEP. The distillation assembly consists of a directly heated distillation container and an
air-cooled condensation bottle. No external cooling system is required.
The OmniPure uses a closed design with a standard storage bottle for collection,
eliminating potential contamination from open-air or transfer bottles. The system is
designed for easy disassembly, cleaning, and exchange of still assemblies. A sliding
bottle rack with a built-in tray provides further ease of use, while protecting against
accidental liquid spillage.
Measured trace element concentrations in HNO3
(by
ICP-MS) before and after a 100°C distillation
| Be |
11.0 |
ND |
| Mg |
11.1 |
0.06 |
| Al |
10.6 |
0.2 |
| Ti |
10.1 |
0.4 |
| V |
10.4 |
ND |
| Cr |
12.5 |
ND |
| Mn |
9.7 |
0.4 |
| Co |
9.4 |
0.01 |
| Ni |
10.1 |
ND |
| Cu |
9.7 |
0.02 |
| Zn |
10.0 |
0.08 |
| As |
10.4 |
ND |
| Se |
9.7 |
ND |
| Sr |
9.3 |
ND |
| Mo |
9.2 |
ND |
| Ag |
9.4 |
0.007 |
| Cd |
9.3 |
ND |
| Sn |
10.8 |
0.5 |
| Sb |
9.7 |
ND |
| Ba |
9.5 |
0.01 |
| Dy |
9.3 |
ND |
| Tl |
9.4 |
ND |
| Pb |
9.5 |
0.004 |
| U |
9.4 |
ND |
ND = Not detected |
Elements were spiked at
10μg/L in the original reagent grade
HNO3. Concentrations for many elements
are below the detection
limit.
Note: Distillation was performed under non-cleanroom conditions.
|