| Element | Wavelength (nm) | Pneumatic Nebulizer (Cross-Flow) |
U-5000AT+ |
| Ag | 328.06 | 2 | 0.3 |
| Al | 396.15 | 9 | 0.5 |
| As | 193.69 | 18 | 2 |
| Be | 313.04 | 0.3 | 0.02 |
| Cd | 214.43 | 2 | 0.2 |
| Co | 228.61 | 5 | 0.3 |
| Cr | 267.71 | 4 | 0.2 |
| Cu | 324.75 | 2 | 0.3 |
| Mo | 202.03 | 8 | 0.3 |
| Ni | 231.60 | 10 | 0.4 |
| Pb | 220.35 | 27 | 1 |
| Sb | 206.83 | 12 | 3 |
| Se | 196.03 | 20 | 1.3 |
| Tl | 190.80 | 22 | 5 |
| V | 292.40 | 2 | 0.2 |
| Zn | 213.85 | 2 | 0.2 |
Notes: Detection limits are based on 3 sigma, with a 10-second integration time. Instrument used was a Thermo Jarrell Ash ICAP 61. |
| [close window] |