ADX-500 Autodilution Accessory

The ADX-500 is designed to work in conjunction with the QuickTrace™ M-7500 Mercury Analyzer adding the ability to automatically dilute and reanalyze samples, which are over the calibration range.

  • No additional bench space required
  • Dual probe system to minimize cross contamination
  • Fully automated
  • Flexible, user defined dilution parameters
  • Software controlled
  • Simple setup

The ADX-500 is designed to work in conjunction with the QuickTrace™ M-7500 Mercury Analyzer adding the ability to automatically dilute and reanalyze samples, which are over the calibration range. The parameters for initiating a dilution are entered by the user. The user can choose to dilute the sample by between 1:2 and 1:10 when intensity is over the high standard. The autosampler positions used for receiving a diluted sample are set by the user. The software also allows the batch analysis to be suspended if all the dilution tubes are used.

The ADX-500 sits on the corner of the autosampler and utilizes a second sampling probe for sample uptake and dispensing of diluent. The second probe minimizes cross contamination because it is only used for diluting over range samples. A 5.0 mL syringe pump is used to precisely perform the dilution with precision of less than 0.05%RSD. The excellent dilution precision, coupled with the low drift QuickTrace™ gives the user worry free automated sample analysis.

Table 1 shows the analytical signal accuracy of the M-7500 with ADX-500 diluting 1/10 and 1/100. The results of dilution accuracy by weight are shown in Table 2. The accuracy of dilution using the ADX-500 is exceptional, with RSD of less than 0.05%.

Table 1: Dilution Signal Accuracy

  10 ppb 10 ppb 10 ppb
Dilution - (1/10 dil) (1/100 dil)
uAbs units 45760 4555 471.5
% recovery - 99.5% 103%
      n=12

Table 2: Dilution Accuracy by Weight

  Total Volume Diluent Sample
True number 10.0 9.00 1.00
Actual average 9.98 8.99 0.99
SD 0.002 0.004 0.005
% RSD 0.02 0.04 0.47
      n=12

Specifications

  • Configuration:  QuickTrace™ M-7500 Automated Mercury Analyzer
  • Software:  Windows based QuickTrace™ Control Software
  • Optional rack sizes:  90 (13mm); 60 (16mm); 40 (20mm); 24 (25mm); 21 (30mm)
  • Computer interface:  RS232
  • Voltage:  100-240 VAC +/-10%, 50/60Hz
  • Unit Dimensions:  Size (in): 12.5 W, 5.0 D, 5.5 H
  • Warranty:  1 year
Features
Variable detection limits
Moderate to high sample throughput
Compatible with various EPA methods
Advanced multi-tasking software